700 CLD / O2
5-04-2017
82
Range Limits
→
→
→
or
→
The analyzer is factory-configured with four physical ranges (1 - 4).
The operator can
change the number of ranges and select a specific full-scale concentration in ppm. From
the Range Setup menu press
to access the Range Limits screen.
To change the Range Limits from the factory settings, use the up or down arrows
to move the highlight to the field you intend to change. Press
to open the field
to change the value in ppm. Press
again to close the field. Press
to save your
changes. To initiate the saved changes, press
, then press
and select new
ranges.
NOTES:
1.
The Range Limit values must be set in ascending order.
2.
The analyzer will not allow any of the range limits to exceed the maximum range
limit on the Range Limits screen.
Example:
Maximum Range Limit 3000.0.
3.
To set a single range, set Range 1 to the desired value and all others to zero.