700 CLD / O2
5-04-2017
49
When the concentration number has stabilized, press
to set the span calibration. The
calibration should now be complete.
If the span calibration was successful, the screen will say ***Saved Current***. If the
calibration was unsuccessful, the screen will say Outside Deviation Limits. See the
Manual Zero Calibration
section for examples of screens showing successful and
unsuccessful calibrations.
If the span calibration is unsuccessful, check the following:
1.
Make sure the correct gas was introduced into the analyzer.
2.
Verify the
Diagnostic Values
while flowing gas.
3.
Check Span Gas Deviations under
Manual Deviations
:
→
→
→
→
→
4.
Check
Maximum Calibration Error
under Calibration Setup:
→
→
→
→
→
After a successful Manual Span Calibration, the analyzer is ready for use.