750 Manual
5-04-2017
155
ADRU: Pressures / Electronic Pressure Control Valve Voltage in Percent
Command
Response
Description
_ADRU_K0
_ADRU_s_z.z_y.y_x.x……
1.
Sample pressure
2.
Air pressure
3.
% of sample EPC volts
4.
% of Ozone EPC Volts
_ADRU_K0_x
_ADRU_s_xpress/x% voltage
Reading of x
ADUF: Flows
Command
Response
Description
_ADUF_K0
_ADUF_s_z.z_y.y
1.
Sample flow
2.
Air flow
_ADUF_K0_x
_ADUF_s_xflow
Flow of x
AEFF: NO
x
Correction Factors
Command
Response
Description
_AEFF_K0
_AEFF_s_z.z_y.y
zz: NO
x
offset
yy: NO
x
correction factor
AEMB: Get Measuring Range
Command
Response
Description
_AEMB_K0
_AEMB_s_Mn
Current range n
AENT: Query Calibration Gas Flow Settings (Pumps or Valves)
Command
Response
Description
_AENT_K0
_AENT_s_x
x: 10 = Pump
11 = Valves