750 Manual
5-04-2017
196
03H Read Floating Point
Modbus Commands Use TCP/IP Port 502. (Do not change TCP/IP port from 7750.)
Register Numbers and Descriptions
Register Number
Contents IEEE
40001
UNDILUTED real-time concentration* dilution ratio / 10000
40003
Real-time concentration (in ppm)
40005
Concentration before linearization and zero / span corrections
40007
Raw detector volts
40009
NO concentration – Switching mode
40011
NO
2
concentration – Switching mode
40013
NO
x
concentration – Switching mode
40017
O
2
concentration (with option)
40019
O
2
concentration before linearization and zero / span corrections
40021
O
2
raw detector volts (with option)
40025
Current range Full-Scale concentration
40027
O
2
range – 25%
40031
Sample pressure
40033
Air pressure
40035
Oven temperature
40037
Converter temperature
40039
Pump temperature
40041
Diode temperature
40043
Cell temperature
40045
Gas dryer temperature (with option)
40047
O
2
temperature (with option)
40049
Sample EPC coil voltage
40051
Air EPC coil voltage
40061
Range 1 offset