700 CLD / O2
5-04-2017
105
The following output signals can be programmed by the operator from the Output
Assignments screen:
RealTime
: In either NO or NO
x
mode the concentration’s output will be a live reading.
NO
: In NO mode the concentration’s output will be a live reading. In NO/NO
x
/NO
2
mode the output will be a read and hold.
NO
2
: The calculated concentration’s output will be updated after each complete cycle in
NO/NO
x
/NO
2
mode.
NO
x
:
In NO
x
mode the output concentration will be a live reading. In NO/NO
x
/NO
2
mode the output will be a read and hold.
SamplePres:
Sample pressure (psig).
AirPres
: Air pressure (psig).
OvenTemp
: Oven temperature (
°
C). (Heated version)
ConvTemp
: Converter temperature (
°
C).
PumpTemp:
Internal pump temperature (
°
C). (Heated version)
DiodeTemp
: Photodiode temperature (
°
C).
CellTemp
: Sample cell temperature (
°
C).
DryerTemp
: Gas temperature of the rear-mounted chiller (
°
C). (Versions with chillers)
Sample EPC
: % of voltage supplied to the Sample electronic proportioning control
valve.
AirEPC:
% of voltage supplied to the Air electronic proportioning control valve.
O2Temp
: Oxygen temperature (
°
C).
O2:
Oxygen concentration.
CaseTemp
: Analyzer’s internal temperature (
°
C).