700 CLD / O2
5-04-2017
101
Press
to set the analyzer in NO
x
mode and enter the Manual Zero Calibration
screen. Introduce zero gas into the rear of the analyzer. Once the concentration stabilizes,
press
to set the Zero Offset. Press
to return to the NO
x
Correction Factors
menu.
Press
to set the analyzer in NO mode and enter the Manual Span Calibration screen.
Introduce NO span gas into the rear of the analyzer. Verify that the span concentration on
the screen matches the Nitric Oxide concentration on the bottle certificate. If the
concentration does not match, press
and change the concentration to match. Once
the concentration stabilizes, press
to set the span. Press
to return to the NO
x
Correction Factors menu.
Press
to set the analyzer in NO
x
mode and enter the Manual Span Calibration
screen. Introduce NO span gas into the rear of the analyzer. Set the concentration on the
screen to match the Total Oxides of Nitrogen concentration from the bottle certificate.
Press
and change the concentration, then press
again to close the field. The
concentration value is temporarily set and will revert back to the original span
concentration. Once the concentration stabilizes, press
to set the NO
x
correction
factor. The correction factors are now set.
Press
to manually edit the NOx Factors.