Auto Calibration Parameters
→
→
→
→
Auto Calibration Parameters allows the operator to select the range, mode and
choose between Zero and All calibration gases (both zero and span).
To access the
AutoCal Parameters screen, press
from the Calibration Setup menu.
To navigate between parameters, use the up or down arrow
to move the
highlight to the field you intend to change. Press
to open the field and change the
parameter. Press
again to close the field after you have made your change.
The first changeable parameter is the Range
to be calibrated. Press
to open the field
and change the range. Then select a range (from 1 – 4) for calibration. To select all
ranges, press 0. Press
to close the field.
The second parameter the operator can change is the Mode. The 700 CLD analyzer can
calibrate in either NO or NOx mode. Press
to open the field and select the mode
using the up/down arrows
. Press
to close the field.