700 CLD / O2
5-04-2017
68
Maximum Verifying Error
→
→
→
→
→
Maximum Verifying Error is the allowable tolerance during the Verifying step of
sequenced calibration.
To access the Maximum Verifying Error screen, press
from
the Deviation Limits menu.
To set the allowable tolerances for different ranges, use the up or down arrow
to move the highlight to the field you intend to change. Press
to open the field to
change the value in %. Press
again to close the field. Press
to save your
changes.