Auxiliary Earth Lead Socket.
Direct Connection Socket.
Auxiliary 10m Earth Lead.
Direct Connection Lead.
Crocodile Clips
and Connection Magnets.
Accessory Tray.
Accessory Tray Retaining Clips.
Battery Compartment.
Batteries.
11
Condition Checks
C.Scope Signal Generators
and Transmitters
The following C.Scope Signal Generators and
Transmitter are covered in this section:
C.Scope grey SGA Signal Generator
C.Scope black SGA Signal Generator
C.Scope grey 1 Watt SGV Signal Generator
C.Scope black SGV Signal Generator
C.Scope MXT Transmitter
C.Scope Signal Generator (Pre-2003)
The following Condition Checks
MUST
be carried out
prior to Performance Checking any C.Scope Signal
Generator or Transmitter with the Function Checker.
These Condition Checks are designed to identify if there
is any mechanical deterioration to the product under
test that may compromise performance or make the
product vulnerable to water or dust ingress.
If the Signal Generator or Transmitter does not pass
all of the Condition Checks then it should be withdrawn
from service immediately and returned to a C.Scope
Authorised Service Centre.