66
Z31823_00_05
Setup and Diagnostics Tool
10.2
Downloading New Firmware
This function is used to download a new firmware version simultaneously to the BMPC/2
Control Unit and Wiring Box. The firmware download typically requires around 10-12
minutes.
i
Notice: Take care when exchanging the Control Unit or Wiring Box, that the same firm-
ware is on the Control Unit and the Wiring Box. The firmware on both units must be the
same!
Normally a BMPC/2 Control Software version works only with a corresponding firmware
version in the Control Unit and the Wiring Box. For example, the firmware
„BMPCAPPL_100201v5.0.H86 has the release date of February 1, 2010, and Version
5.0. When the BMPC/2 is started, it checks if the version V5.0 corresponds to the inter-
nal stored version Vx.y. If the major release value („x“) is equal, the software will start,
otherwise a warning window will appear and the software will not start.
After selecting this function, browse for the actual firmware file which should be stored
before in the directory:
c:\program files\bruker\bmpc2\Firmware\bmpc*.h86
After selecting the current version, you can start the download by pressing the button
Start Download
.
Figure 10.2 Setup Tool - Download
Download in progress
Download finished
Summary of Contents for BMPC/2
Page 8: ...8 Z31823_00_05 Table of Contents...
Page 14: ...14 Z31823_00_05 Safety...
Page 20: ...20 Z31823_00_05 Principles of Operation...
Page 42: ...42 Z31823_00_05 Alarm System...
Page 62: ...62 Z31823_00_05 Hardware Overview...
Page 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Page 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Page 122: ...122 Z31823_00_05 Technical Data...
Page 124: ...124 Z31823_00_05 Contact...
Page 126: ...126 Z31823_00_05 Warning Signs...
Page 130: ...130 Z31823_00_05 Figures...
Page 132: ...132 Z31823_00_05 Tables...
Page 135: ......