Alarm System
6.2
Alarm Configuration
All channels that have the option „Alarm Trigger“ checked can trigger an alarm cycle.
The alarm trigger option can be configured using the function from the menu
Tools -
Options - Sensor Channels
.(for additional information see
).
An alarm will also be triggered if the communication between the industrial PC and the
BMPC/2 Control Unit (RS232) fails. This may occur, for example, when a power failure
occurs, which lasts longer then the backup time of the external UPS, or which is not pro-
tected by the external UPS. The PC then will be powered for 30 more minutes using the
internal UPS, whereas an alarm cycle can be initiated during this time.
6.3
Alarm Sequence
When the system detects a deviation above or below a limit value, then an alarm cycle
will be started.
1. After 30 seconds a check will be made to see if the deviation still exists. When the
deviation still exists, the alarm cycle will continue (start calling), otherwise, the alarm
cycle will be terminated.
No Deviation
: Alarm cycle will be terminated
2. When the cycle continues the people from the
(see
) are called sequentially. To learn what you have to do when
you receive an alarm call, please see
Acknowledgement
: Alarm calls stopped. When the person who has assumed
responsibility for the system (by acknowledging the alarm call) switches the sys-
tem on site to „OFF“ mode, a confirmation fax will be sent to all alarm fax recipi-
ents to show that the magnet system is under control.
3. When no acknowledgement takes place the
Acknowledgement
: Alarm calls stopped. When the person who has assumed
responsibility for the system (by acknowledging the alarm call) switches the sys-
tem on site to „OFF“ mode, a confirmation fax will be sent to all alarm fax recipi-
ents to show that the magnet system is under control.
4. When the fax group is enabled, a fax will be sent to the entire
5. When this option is enabled, an e-mail will be sent to the entire
.
Summary of Contents for BMPC/2
Page 8: ...8 Z31823_00_05 Table of Contents...
Page 14: ...14 Z31823_00_05 Safety...
Page 20: ...20 Z31823_00_05 Principles of Operation...
Page 42: ...42 Z31823_00_05 Alarm System...
Page 62: ...62 Z31823_00_05 Hardware Overview...
Page 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Page 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Page 122: ...122 Z31823_00_05 Technical Data...
Page 124: ...124 Z31823_00_05 Contact...
Page 126: ...126 Z31823_00_05 Warning Signs...
Page 130: ...130 Z31823_00_05 Figures...
Page 132: ...132 Z31823_00_05 Tables...
Page 135: ......