Principles of Operation
4.4.1
P1 Pressure Monitoring
The P1 pressure is continuously monitored against a pre-specified upper and lower limit
and the following actions taken if one of these limits is reached:
4.4.2
T1/T2 Temperature Monitoring
The T1 and T2 temperatures are continuously monitored against a pre-specified lower
limit and the following actions taken if the limit is reached:
4.4.3
Fl Helium Flow Monitoring
The Fl helium flow value is continuously monitored against a pre-specified lower limit
and the following actions taken if the limit is reached:
Precondition
Condition
Action
P1 pressure is
lower
than
P1
low warning
limit
A warning is
triggered
.
Software in Monitor &
Alarm mode
P1 pressure is
lower
than
the P1
low alarm
limit
Alarm cycle is
activated
P1 pressure is
higher
than
the P1
high warning
limit
Second pump is started and
a warning is
triggered
Software in Monitor &
Alarm mode
P1 pressure is
higher
than
the P1
high alarm
limit
Second pump is started and
the alarm cycle is
activated
Table 4.1 P1 Pressure Monitoring
Precondition
Condition
Action
T1 or T2 temperature is
lower
than the T1/T2
warn-
ing
value
Second pump is started and
a warning is
triggered
Software in Monitor &
Alarm mode
T1 or T2 temperature is
lower
than the T1/T2
alarm
value
Second pump is started and
the alarm cycle is
activated
Table 4.2 T1/T2 Temperature Monitoring
Precondition
Condition
Action
Fl value is
lower
than the
flow
warning
value
Second pump is started and
a warning is
triggered
Software in Monitor &
Alarm mode
Fl value is
lower
than the
flow
alarm
value
Second pump is started and
the alarm cycle is
activated
Table 4.3 F1 Helium Flow Monitoring
Summary of Contents for BMPC/2
Page 8: ...8 Z31823_00_05 Table of Contents...
Page 14: ...14 Z31823_00_05 Safety...
Page 20: ...20 Z31823_00_05 Principles of Operation...
Page 42: ...42 Z31823_00_05 Alarm System...
Page 62: ...62 Z31823_00_05 Hardware Overview...
Page 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Page 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Page 122: ...122 Z31823_00_05 Technical Data...
Page 124: ...124 Z31823_00_05 Contact...
Page 126: ...126 Z31823_00_05 Warning Signs...
Page 130: ...130 Z31823_00_05 Figures...
Page 132: ...132 Z31823_00_05 Tables...
Page 135: ......