102
Z31823_00_05
Configuration (administrator)
•
P2 Heater Set Point: This is the target value of the automatic heater control. This
value must be at least 1 mBar lower as the P2 APD target value.
•
P2 APD: This is the target value for the P2 absolute pressure regulator. The abso-
lute pressure regulator ensures that the magnet remains at a constant pressure
even when working in the over pressure range.
•
Heater Power H1: This field allows you to enter the H1 heater heat value for heater
operation in manual mode.
•
Heater Power H2: This field allows you to enter the H2 heater heat value for heater
operation in manual mode.
•
Heater Resistor H1: This field is used to enter the resistance value for the H1
heater.
•
Heater Resistor H2: This field is used to enter the resistance value for the H2
heater.
Press the
Write Values
button to write the values to the unit.
CAUTION
Heater and APD Settings
Changes in these adjustments may cause low pressure in the magnet dewar. This
could lead to a dangerous icing of the magnet.
Changes should only be made by experienced personnel that understand the
implications of these changes.
Summary of Contents for BMPC/2
Page 8: ...8 Z31823_00_05 Table of Contents...
Page 14: ...14 Z31823_00_05 Safety...
Page 20: ...20 Z31823_00_05 Principles of Operation...
Page 42: ...42 Z31823_00_05 Alarm System...
Page 62: ...62 Z31823_00_05 Hardware Overview...
Page 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Page 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Page 122: ...122 Z31823_00_05 Technical Data...
Page 124: ...124 Z31823_00_05 Contact...
Page 126: ...126 Z31823_00_05 Warning Signs...
Page 130: ...130 Z31823_00_05 Figures...
Page 132: ...132 Z31823_00_05 Tables...
Page 135: ......