He Fill Mode
The He Fill mode is necessary because the T1 and/or T2 parameters may briefly fall
below the alarm threshold during the helium refill process. In order to avoid an alarm
from being triggered the He Fill mode can be used to suppress the T1/T2 alarm for up to
six hours.
The P2 pressure control heater is also disabled for one hour, as this is not efficient when
the magnet is open and thus would only lead to unnecessary helium consumption.
When the He Fill mode is activated it is indicated in the status bar together with the
length of fill time remaining.
Figure 5.16 He Fill Mode 292 Minutes
The suppressed alarms are represented in the Present Values/Event display as indi-
cated in the figure below.
Figure 5.17 Alarm Temporarily Disabled
Once the helium refill is finished and the remaining time was expired, the program will
automatically leave He Fill mode and return to monitoring mode.
For information on configuration the He Fill mode read
"He-Refill / He Measurement" on
Alarm temporarily
disabled
Summary of Contents for BMPC/2
Page 8: ...8 Z31823_00_05 Table of Contents...
Page 14: ...14 Z31823_00_05 Safety...
Page 20: ...20 Z31823_00_05 Principles of Operation...
Page 42: ...42 Z31823_00_05 Alarm System...
Page 62: ...62 Z31823_00_05 Hardware Overview...
Page 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Page 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Page 122: ...122 Z31823_00_05 Technical Data...
Page 124: ...124 Z31823_00_05 Contact...
Page 126: ...126 Z31823_00_05 Warning Signs...
Page 130: ...130 Z31823_00_05 Figures...
Page 132: ...132 Z31823_00_05 Tables...
Page 135: ......