85
Z31823_00_05
Troubleshooting
11.8.2
Exchanging the GS1 or GS2 Power Supply Fuses
There are two power supplies in the BMPC line control unit. Each of these power sup-
plies has its own safety fuse.
The following fuse type is required for the power supplies:
•
Cartridge fuse 5 x 20 mm 0.8 A time lag. The Bruker part number is
2250
.
The fuse for power supply GS1 is FU1 and the fuse for power supply GS2 is FU2. See
the previous section (
) for instructions on how to exchange the fuse.
11.8.3
Exchanging the Control Unit Fuse
The BMPC/2 control unit utilizes two safety fuses.
The following fuse type is required for the control unit:
Cartridge fuse 5 x 20 mm 0.8 A time lag. The Bruker part number is
2250
.
To exchange a fuse in the control unit:
1. Remove the upper rear cover. To access the back of the control unit, unscrew the
left and right side screws securing the line control unit using a cross-tip screwdriver
and tilt the unit outwards.
Figure 11.9 Tilt Open the Line Control Unit
2. Remove the power supply plug from the control unit and open the fuse holder as
shown in the figure below.
3. Remove the fuse insert and exchange both safety fuses.
4. Reassemble the unit in reverse order.
Unscrew
the left
Tilt open the
Line Control Unit
and right
screws
Summary of Contents for BMPC/2
Page 8: ...8 Z31823_00_05 Table of Contents...
Page 14: ...14 Z31823_00_05 Safety...
Page 20: ...20 Z31823_00_05 Principles of Operation...
Page 42: ...42 Z31823_00_05 Alarm System...
Page 62: ...62 Z31823_00_05 Hardware Overview...
Page 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Page 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Page 122: ...122 Z31823_00_05 Technical Data...
Page 124: ...124 Z31823_00_05 Contact...
Page 126: ...126 Z31823_00_05 Warning Signs...
Page 130: ...130 Z31823_00_05 Figures...
Page 132: ...132 Z31823_00_05 Tables...
Page 135: ......