38
Z31823_00_05
Software
5.5.3
Manual Start of the Pump Test
It is also possible to check the function of the spare pump manually. To start the pump
test manually:
1. Select the
OFF
mode in the menu bar.
2. Select
Tools
-
Options
-
Pump Control
-
Pump Test
.
3. Ensure that the checkbox „Enable Automatic Pump Test“ is checked.
4. Press the „Start pump test now“ button (
1).
5. The pump test will proceed as described in the automatic test. A status bar and text
output will show the status of the pump test cycle.
During the test the P1 field will display the P1 pressure and chart the values over
time. Likewise the helium flow will be displayed in the He Flow field and the values
over time will be charted. In both cases the x-axis will be adjusted automatically
Figure 5.20 Tools - Options - Control Unit - Pump Test
i
If the automatic pump test is disabled, a faulty backup pump will not be recognized. Thus
when the main pump fails there will be no backup up pump which could lead to instabili-
ties of the magnet. This will also trigger an alarm cycle.
1
Summary of Contents for BMPC/2
Page 8: ...8 Z31823_00_05 Table of Contents...
Page 14: ...14 Z31823_00_05 Safety...
Page 20: ...20 Z31823_00_05 Principles of Operation...
Page 42: ...42 Z31823_00_05 Alarm System...
Page 62: ...62 Z31823_00_05 Hardware Overview...
Page 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Page 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Page 122: ...122 Z31823_00_05 Technical Data...
Page 124: ...124 Z31823_00_05 Contact...
Page 126: ...126 Z31823_00_05 Warning Signs...
Page 130: ...130 Z31823_00_05 Figures...
Page 132: ...132 Z31823_00_05 Tables...
Page 135: ......