Configuration (administrator)
In the setup category LIMIT VALUES (see chapter "Limit Values") you will find among
others the following two entries used to setup the limit values for use with the HLMU:
- Helium Level (HLMU)
- Nitrogen Level (HLMU)
It is recommended to use the given default parameters (Normal = 70, Warning = 60,
Alarm = 50).
Note: If you are using the HLMU for helium (nitrogen) level measurements, the channel
"Helium Level" that is normally used, must be disabled (see
), and the two channels Helium Level (HLMU) and Nitrogen Level (HLMU) must be
enabled.
When you now start the Monitoring Mode/Monitoring & Alarm Mode the system will first
try to establish a connection to the HLMU. If it fails, the system will not start and an entry
will be logged in the Event Log. When the connection is successful, the system will begin
continuously reading the level(s) values from the HLMU.