Software
5.5.2
Configuring the Automatic Pump Test
The automatic pump test can be configured to start at a pre-designated date and time.
To configure the automatic pump test:
1. Select the
OFF
mode in the menu bar.
2. Select
Tools
-
Options
-
Pump Control
-
Pump Test
3. Check the „Enable Automatic Pump Test“ checkbox (
4. Select the date and time for the next test (
5. Select the interval for the test (
3). A pump test once a month (every 4
weeks) is recommended.
6. Press
OK
to apply the configuration.
Figure 5.18 Tools - Options - Control Unit - Pump Test - Configuration
After the automatic pump test is configured it may be enabled or disabled using the
checkbox listed in step 3 above. The date and time of the next scheduled test is shown
in the event log window (display mode „Present Values/Events“) every time you start the
Monitoring & Alarm mode (
1). If the pump test is disabled this is also shown
(
2).