53
Z31823_00_05
Hardware Overview
Fuse
Purpose of Fuse
Value
Q1
Main fuse for:
Power Supply GS 1/2, Pump 1/2, Valve 1/2/3,
Warning Lamp, Industrial PC, Control Unit,
Fan 1/2/3 (Ventilator 1/2/3)
16A
Q2
Cryo Platform Out (X4:26)
6A
Q3
Console Out (X3:4)
30A (ECL00: 32)
Q4
Valve 1 (X4:17)
6A
Q5
Valve 2 (X4:18)
6A
Q6
Valve 3 (X4:19)
6A
Q7
Warning Lamp (X4:20)
4A
Q8
Cryo-PSU (X4:27)
16A
Q9
Industrial PC/Monitoring System
(X4:21)
4A
Q10
Control Unit (X4:22)
6A
FU1, 2
Main Fuses - Main Input
50A
P/N: 87570
FU3
Power Supply - GS1
0.4A/T
BRUKER P/N: 2247
FU4
Power Supply - GS2
0.4A/T
BRUKER P/N: 2247
FU5
Fan 1 (Ventilator 1)
0.1A/T
BRUKER P/N: 2241
FU6
Fan 2 (Ventilator 2)
0.1A/T
BRUKER P/N: 2241
FU7
Fan 3 (Ventilator 3)
0.1A/T
BRUKER P/N: 2241
Table 8.2 Fuse Overview