13 Technical Data
13.1
Voltage Supply Requirements
Input Voltage:
208 -230 VAC (fluctuations of less than + 10%)
Input Frequency Range: 50 - 60 Hz
Pollution Scale:
2
Over Voltage Category: 2
13.2
Site Conditions
The BMPC/2 should only be used:
•
In a standard laboratory environment.
•
Up to a maximum elevation of 2000 meters.
•
At a temperature between 15-35 °C.
•
A relative humidity of a maximum of 80% for temperatures up to 31°C, linearly
decreasing to 67% at a temperature of 35 °C.
13.3
Noise Level
The measured noise level for the BMPC/2 is less than 45 dBA.
13.4
Power Input Safety Fuse
The power input safety fuse for the BMPC/2 is a 50A dual-element time delay fuse Class
J 600V. The Bruker part number for this fuse is
87570
. Two fuses of this type are used in
the unit.