Troubleshooting
11.7
Turning on the NMR Console After a Power Failure
When the NMR console switches off due to a power failure there are two different ways
to restart it: when the power is restored:
•
Restart the console through the software.
•
Restart the console using the mains switch.
11.7.1
Restarting the NMR Console Through the Software
In the menu
Tools
you will find the options
Enable NMR Console
and
Disable NMR Con-
sole
. When you are logged in as Administrator you can use these functions to turn the
NMR console on or off.
Figure 11.6 Restarting the NMR Console Through Software
11.7.2
Restarting the NMR Console Using the Mains Switch
If you cannot switch on the console using the software, you can alternatively use the S4
switch. The S4 switch is found behind the rear cover, roughly in the middle. As soon as
you turn on this switch the KM4 protection will be activated and the console will be provi-
ded with power.
Figure 11.7 Restarting the NMR Console Using the Main Switch
Turn-on the Console
Turn-off the Console
The S4 switch is used
to turn on the
NMR console.
Summary of Contents for BMPC/2
Page 8: ...8 Z31823_00_05 Table of Contents...
Page 14: ...14 Z31823_00_05 Safety...
Page 20: ...20 Z31823_00_05 Principles of Operation...
Page 42: ...42 Z31823_00_05 Alarm System...
Page 62: ...62 Z31823_00_05 Hardware Overview...
Page 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Page 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Page 122: ...122 Z31823_00_05 Technical Data...
Page 124: ...124 Z31823_00_05 Contact...
Page 126: ...126 Z31823_00_05 Warning Signs...
Page 130: ...130 Z31823_00_05 Figures...
Page 132: ...132 Z31823_00_05 Tables...
Page 135: ......