7
en
2.3.3
Inserting the implant in the implant
bed
Picking up the implant with the insertion instru-
ment
►
Select an implant according to the size of the trial
implant.
►
Remove the implant with protective packaging
from the sterile packaging.
►
Mount the implant
a
in the protective packaging
onto the dihedron
7
of the insertion instrument
A
,
see Fig. 7. When doing so, pay attention to the ori-
entation of the implant (orientation according to
the marking on the insertion instrument).
Fig. 7
Coupling the implant to the insertion instru-
ment
Legend
a
Implant
►
Turn the switch
2
to the right (direction ), see
Fig. 3.
►
Hold the implant at 0° position, see Fig. 8.
Fig. 8
Implant at 0° position
►
Turn the rotary nut
3
on the insertion instrument
A
clockwise (:“tighten” direction) until the trial
implant is clamped, see Fig. 4.
►
Remove the protective packaging from the implant.
►
Check manually that the implant is securely in
place.
Filling the implant with bone material (optional)
►
Place implant
a
in filling holder
E
, see Fig. 9. When
doing so, observe the implant length (marking on
the filling holder: L26/L30/L34).
Fig. 9
Implant in filling holder
►
Compress the bone grafts or bone replacement
material in the inner window of the implant
a
using
the pestle provided in the set. When doing so,
ensure that the pestle does not damage the surface
of the implant.
Inserting the implant
►
Insert the implant
a
with the insertion
instrument
A
straight (0° position) into the pre-
pared implant bed. To do so, carefully hit with a
hammer on the impact surface of the rotary nut
3
.
Fig. 10 Inserting the implant into the implant bed
a
7
a
A
a
7
A
a
E
A
a
a
A