14
Aesculap
®
High frequency surgical device GN060
5.5
Wipe disinfection for electrical devices without sterilization
RT: Room
temperature
Phase I
►
Remove any visible residues with a disposable disinfectant wipe.
►
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
►
Observe the specified application time (1 min minimum).
5.6
Inspection, maintenance and checks
►
Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
►
Inspect all cables, and in particular electrode cables, for any damage to
their insulation.
►
Set aside the product if it is damaged.
►
Perform a safety inspection annually, see Maintenance.
5.7
Storage and transport
►
Transport the product only in its original box.
►
Observe storage and transport conditions, see Ambient conditions.
6.
Maintenance
The HF device GN060 does not require maintenance.
Note
Maintenance must only be carried out by authorized personnel. If neces-
sary, the circuit diagrams and the service manual, which contains all the
necessary documents, are provided for this.
6.1
Safety inspection
A safety inspection must be carried out annually.
The person in charge of the inspection will document the inspection find-
ings and measurement values according to the printed inspection proto-
col.
►
Only have the product and its accessories inspected by persons who
possess the requisite training, knowledge and experience and who are
authorized to do so.
►
If there are significant deviations from the inspection protocol values,
or if the prescribed maximum values are exceeded: Send in device.
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
Phase
Step
T
[°C/°F]
t
[min]
Conc.
[%]
Water quality
Chemical
I
Wipe disinfection
RT
≥
1
-
-
Meliseptol HBV wipes 50 % Propan-1-ol
CAUTION
Damage to the product caused by switch-on to
early after storage/transport at temperature below
+10 °C (50° F)!
►
Allow the HF device to acclimate at room tem-
perature for about 3 hours.
Summary of Contents for Aesculap GN060
Page 2: ......
Page 3: ...1 2 3 9 5 4 6 7 8 12 12 10 13 11 14 15 14 16 18 19 12 12 20 13 13 17 21 22 23...
Page 150: ...148 Aesculap GN060 1 GN060 TA022130 Aesculap IEC VDE 1 1 IEC VDE...
Page 151: ...149...
Page 159: ...157 F2 30 F3 30 F3 20 20 10 4 5 CPU RAM EPROM 10 RAM RAM EEPROM...
Page 161: ...159 5 3 5 4 5 5 I T C F t I 1 Meliseptol HBV 50 1...
Page 162: ...160 Aesculap GN060 5 6 5 7 6 GN060 6 1 B Braun Aesculap 10 C 3...
Page 167: ...165 10 1 700 3000 11 B Braun Aesculap Extranet PDF...