8
Aesculap
®
High frequency surgical device GN060
3.3
Moving stacks of units
►
Always lift stacks from underneath the bottommost unit.
3.4
Dismantling stacks of devices
Dismounting a stack of products fitted with stacking devices:
Fig. 5
Dismantling stacks of devices
Legend
B
Locking screw
►
Apply a screwdriver on locking screw
B
and turn it counterclockwise to
the positive stop.
►
Remove the topmost device.
►
Screw back on the feet of the device, either by hand or using a screw-
driver.
3.5
Presetting the correct voltage
The mains voltage must correspond to the voltage indicated on the type
plate at the back of the unit.
The correct voltage for the control unit is indicated in the window located
at the back of the device. The standard voltage for the unit in Europe is:
230 V.
►
If the indicated voltage does not correspond to the actual mains volt-
age, change the setting at the rear panel of the power unit:
– Release clip
B
using a suitable screwdriver
A
and remove the fuse
holder
25
.
– Pull out the voltage selection element
24
under the fuse holder.
– Reinsert the voltage switching element so that the desired
voltage
C
is legible.
– Reinsert the fuse holder
25
.
Fig. 6
Device connection with fuse holder and voltage selection ele-
ment
Legend
23
Device connector
24
Voltage selection element
25
Fuse holder
A
Screwdriver
B
Clip
C
Voltage switching element imprint
DANGER
Risk of fatal injury from electric shock!
►
Unplug the device before replacing the fuse sets.
B
Voltage range
Imprint C
100 V to 120 V (+/- 10 %)
110
220 V to 240 V (+/- 10 %)
230
23
A
25
24
C
B
Summary of Contents for Aesculap GN060
Page 2: ......
Page 3: ...1 2 3 9 5 4 6 7 8 12 12 10 13 11 14 15 14 16 18 19 12 12 20 13 13 17 21 22 23...
Page 150: ...148 Aesculap GN060 1 GN060 TA022130 Aesculap IEC VDE 1 1 IEC VDE...
Page 151: ...149...
Page 159: ...157 F2 30 F3 30 F3 20 20 10 4 5 CPU RAM EPROM 10 RAM RAM EEPROM...
Page 161: ...159 5 3 5 4 5 5 I T C F t I 1 Meliseptol HBV 50 1...
Page 162: ...160 Aesculap GN060 5 6 5 7 6 GN060 6 1 B Braun Aesculap 10 C 3...
Page 167: ...165 10 1 700 3000 11 B Braun Aesculap Extranet PDF...