Power Analysis
198
19.11
Power Efficiency
Power efficiency analysis can test the overall efficiency of the power supply by measuring the output power and input
power. This analysis is only supported on 4-channel models because all of the input voltage, input current, output voltage,
and output current are necessary for the measurement.
19.11.1
Power Efficiency Input Setup
1.
Touch
Input Setup
to display the signal menu shown in figure
2.
Touch
Input Voltage
to set the input voltage source.
3.
Touch
Input Current
to set the input current source.
4.
Touch
Output Voltage
to set the output voltage source.
5.
Touch
Output Current
to set the output current source
6.
Touch
Duration
and turn the
Universal Knob
or use the virtual numeric
keypad to set the duration time.
–
the range is 5 ns 20 s.
–
The oscilloscope will set the timebase according to the value (duration /
10)
7.
Touch
Connection Guide
to display a diagram illustrating the procedure
setup. See figure
Figure 19.26
Input
Setup Menu
Figure 19.25
Connection Guide Power Efficiency