Serial Trigger and Decode
113
10.9
MIL-STD-1553B Trigger and Serial Decode
This section covers triggering and decoding MIL-STD-1553B signals. Please read the
following for more details:
MIL-STD-1553B Signal Configuration
, and
Before configuring the signals settings connect the MIL-STD-1553B signal to the
oscilloscope.
10.9.1
MIL-STD-1553B Signal Configuration
To select the decoding
Bus Protocol
:
1.
Press the Decode key or use the touchscreen controls to enter the
Decode Menu
.
2.
Use the touchscreen to enable
Bus Operation
.
3.
Use the touchscreen to select
Bus Protocol
.
4.
will appear with the addition of the
Manchester
protocol.
5.
Select the
MIL-STD-1553B
bus protocol.
Figure 10.45
MIL-STD-1553B Decode
Protocol Signals
To set the source and threshold of the
MIL-STD-1553B
signal use the touchscreen to
select
1.
The
MIL-STD-1553B
menu will be displayed. (Figure
2.
Use the touchscreen to select
Source
and select the source the signal is connected to.
3.
Use the touchscreen to select
Upper Threshold
and input the desired threshold
–
The threshold voltage level is for decoding, and it will be regarded as the trigger
voltage level when the trigger type is set to serial.
–
The threshold level line only appears when adjusting the threshold level.
4.
Use the touchscreen to select
Lower Threshold
and input the desired threshold
–
The threshold voltage level is for decoding, and it will be regarded as the trigger
voltage level when the trigger type is set to serial.
–
The threshold level line only appears when adjusting the threshold level.
5.
Touch
option to return previous menu.
Figure 10.46
MIL-STD-1553B Signal Menu