Serial Trigger and Decode
108
10.8
I2S Trigger and Serial Decode
This section covers triggering and decoding I2S signals. Please read the following for more
details:
and
Before configuring the signals settings connect the I2S signal to the oscilloscope.
10.8.1
I2S Signal Configuration
To select the decoding
Bus Protocol
:
1.
Press the Decode key or use the touchscreen controls to enter the
Decode Menu
.
2.
Use the touchscreen to enable
Bus Operation
.
3.
Use the touchscreen to select
Bus Protocol
.
4.
will appear with the addition of the
Manchester
protocol.
5.
Select the
I2S
bus protocol.
Protocol Signals
To set the source and threshold of the
I2S
signal use the touchscreen to select
1.
The
I2S SIGNAL
menu will be displayed. (Figure
BCLK
2.
Use the touchscreen to select
BCLK
, the
Source Setting
window will appear.
•
Select the channel that will provide the bit clock signal.
•
Use the touchscreen to select
Threshold
and input the desired threshold.
–
The threshold voltage level is for decoding, and it will be regarded as the trig-
ger voltage level when the trigger type is set to serial.
–
The threshold level line only appears when adjusting the threshold level.
•
Use the touch screen to select either
Rising
or
Falling
edge.
–
Rising
: Data latched on the rising edge of the clock.
–
Falling
: Data latched on the falling edge of the clock.
Figure 10.40
I2S Signal Menu