Measure
The oscilloscope provides measurements of 36 waveform parameters and the statistics. It contains voltage, time, and
delay parameters. The 2560B series can measure multiple channels at the same time, displaying up to 5 parameter
measurements with statistics while in the M1 display mode and up to 12 parameters in the M2 mode. To view more
parameters on a specified channel, the
Simple
mode can be employed. For measuring waveform of interest in a time
gate, the
Gate
function is suggested.
Some parameter measurements (such as the mean) are generated by all the data in a frame. Some parameter
measurements (such as period) accumulate all measurements in a frame, but the displayed value is always the first
value. The distribution of multiple parameters in one frame is displayed in the statistics function.
Vertical, Horizontal, and Miscellaneous measurement can be found in the
Type
option.
12.1
Measurement Modes
1.
Press the
Measure
key to open the
Measure
menu.
2.
Touch the
Measure
option to toggle the measure function on or off.
3.
Touch the
Mode
option to toggle the measure mode between
Simple
and
Advanced
.
•
Simple
: Displays the specified basic measurement parameters of the selected
channel.
Figure 12.1
Simple Mode
If the parameter is unable to be measured, it will display as
demonstrates this for the CCJ measurement
Figure 12.2
Measure Menu
Note: