50
ADVANCED USE MANUAL DD700
2.8.4
Weighing request from key and transmission from serial command
After loading the weight onto the weighing platform, wait
for it to indicate that the weight is stable and press
; the terminal will store the data.
The calculator can request the weight data from the terminal through the following command:
MP<CR>(MPb<CR>)
The weighing terminal responds:
with one of the strings described in
par. 2.8.2
;
NP<CR><LF>
no weight has been stored;
The data can be requested several times with the MP<CR> command.
At the end of the operation the calculator must send the command to enable the terminal for a new
weighing operation:
MP<CR>
(MPb<CR>)
2.8.5
Weighing and transmission request from serial command
After loading the weight onto the weighing platform, send the storage command through the calculator:
MP<CR>(MPb<CR>)
When the weight validity conditions are reached, the terminal stores the data.
The external processor then sends the data request command:
MP<CR>(MPb<CR>)
The terminal responds with one of the strings described in
par.2.8.2
.
2.8.6
MP and MC remote commands with checksum
MP and MC remote commands change their structure for the inclusion of checksum characters. Their
format changes to:
MP 1 D <CR>
MC 0 E <CR>