Flexible time performs calibrations at various intervals. The system uses an
algorithm to determine the number of minutes between calibrations. This algorithm
is based on sensor status and the change in drift values from previous calibrations.
Calibrations are performed at intervals necessary to avoid excessive drift to the
sensor that is experiencing the fastest rate of drift. The default time for the
minimum time between calibrations is 10 minutes. You can define the minimum
time between intervals for one-point calibrations in setup. The minimum time for
two-point calibrations is four times the value entered for one-point calibrations.
Refer to Selecting Calibration Frequency and Automatic Repeat in Section 5 for
more information about defining the time interval for automatic calibrations.
The system indicates that an automatic calibration is pending 5 minutes before the
calibration is scheduled and continues to display the time remaining until the
calibration starts. The sample door closes when a calibration starts. The timing bar
indicates the approximate time to calibration completion. During a calibration, you
cannot access the Menu mode.
You can defer or interrupt calibrations to analyze samples. You can defer automatic
calibrations for up to 30 minutes. After 30 minutes, the system is not in calibration.
The system then starts the required calibration automatically. The system does not
accept samples until the required calibration finishes.
In addition to the regularly scheduled one-and two-point calibrations, the 860
system has a metabolite recal option. The metabolite recal option allows the system
to perform an automatic one-point metabolite calibration to minimize within run
drift during rapid sampling. Metabolite recal is particularly useful in controlling
within run drift for newly installed glucose biosensors. Turning off the metabolite
recal option will increase sample throughput time. With metabolite recal off,
glucose is calibrated only at the regularly scheduled one and two-point
calibrations.
Metabolite recal also decreases within run glucose drift when running many
replicates of QC material. Turning off the recal option does not significantly affect
QC sample results, as most QC samples are limited to 6 replicates in a run.
Individual laboratories should determine if glucose QC recovery is acceptable with
metabolite recal off.
Metabolite recal calibrates the glucose biosensor under two sets of criteria in a
cyclical manner. First, the system uses an algorithm to determine the frequency of
samples. If two consecutive QC samples or a combination of three QC and patient
samples are analyzed at an average rate of 4 minutes or less between samples, the
system begins a metabolite one-point calibration at the completion of the analysis.
If a regularly scheduled one- or two-point calibration is due, the system performs
the scheduled calibration instead.
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......