30. Verify that the sensors are installed in the following order:
31. Close the measurement module door.
32. Install the reagent pump tubing:
a. Disconnect the right side of the tubing from right positions 1 and 2.
b. Remove the tubing cuff from the right side of the platen.
c. Place the tubing over the top of the roller cage.
d. Hold the right tubing cuff, turn the roller cage clockwise, and work the new
tubing between the platen and roller cage. Do not stretch the tubing.
It is normal for the pump tubing to fit tightly.
e. Place the right tubing cuff under the right side of the platen.
f. Connect the right end of the large tubing to position 1.
g. Connect the right end of the small tubing to position 2.
33. Install the sample pump and waste pump tubing:
a. Disconnect the right side of the sample pump tubing.
b. Place the tubing over the top of the roller cage.
c. Hold the right tubing cuff, rotate the roller cage clockwise, and gently work
the new tubing between the platen and roller cage. Do not stretch the
tubing.
It is normal for the pump tubing to fit tightly.
d. Place the right tubing cuff under the right side of the platen.
e. Connect the right tubing to the manifold at position 4.
f. Repeat steps a through d for the waste pump tubing.
g. Connect the right tubing to the manifold at position 5.
34. Install the printer paper:
a. Lift the printer cover up.
b. Place the paper roll in the cavity with the paper unrolling from
the bottom.
c. Lift the printer lever up.
d. Push the paper under the platen until it comes out of the other side in front.
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......