For testing syringe, capillary, microsyringe, and microcapillary modes, blood was
collected in heparinized vacuum tubes. It was tonometered at 37.0
C to each of
three levels to prepare samples for pH analysis, and five levels to prepare samples
for pCO
2
and pO
2
analysis. Blood was spiked/diluted to each of three levels to
prepare samples for sodium, potassium, calcium, chloride glucose and lactate
analysis. Multiple runs were made using these samples on four 860 systems, except
for lactate. For lactate, multiple runs were made on six 860 systems. The
experimental protocol called for three replicates of each level in each run.
For testing the expired gas mode, 10 mL of tonometry gas were drawn into a
12 mL syringe and aspirated into the 860 system for analysis. Multiple runs were
made using five levels of expired gas on four 860 systems. The protocol called for
three replicates of each level in each run.
Table E-31 through Table E-44 summarize the results of the 860 system whole
blood, expired gas, and electrolyte recovery and precision testing.
*() $!' !" "#+
$
"!$
& #
$!'
7.21
Syringe
48
0.002
7.213
7.214
100.0
Capillary
48
0.004
7.202
7.202
100.0
Microcapillary
48
0.003
7.205
7.205
100.0
Microsyringe
48
0.002
7.204
7.205
100.0
pH/Lyte
48
0.003
7.207
7.208
100.0
7.37
Syringe
48
0.002
7.373
7.369
100.1
Capillary
44
0.002
7.376
7.375
100.0
Microcapillary
48
0.006
7.376
7.376
100.0
Microsyringe
48
0.002
7.371
7.370
100.0
pH/Lyte
48
0.002
7.361
7.363
100.0
7.52
Syringe
48
0.003
7.544
7.546
100.0
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......