Sensor drift is beyond predefined limits during a one-point or a two-point calibration.
Qualifiers:
pH
K
+
Ca
++
Cl
–
Na
+
If D codes also exist for reagent problems such as D23, D24, or D29,
correct these problems first, then correct this D code.
If more than one sensor has excessive drift, check the reference sensor for KCl
leaks as described in Cleaning the Reference Sensor and Removing Bubbles in
Section 3.
Check the expiration dates and the levels of the reagents. Ensure that the
reagent bottles are installed properly and in the correct location. If reagents
have expired, are low, or are incorrectly installed, replace the affected reagents,
prime the system, and perform a two-point calibration.
Deproteinize and condition as described in Deproteinizing the Sample Path and
Conditioning the Sensors in Section 3.
Check the solution level in the affected measurement sensor and look for
bubbles. If the solution level is low, refill the sensor as described in Filling the
Measurement Sensors in Section 3.
Check the solution level in the reference sensor and look for salt in the vent
hole, bubbles in the sensor, or leaks. If the solution level is low or salt, bubbles,
or leaks are present, clean and refill the reference sensor as described in
Cleaning the Reference Sensor and Removing Bubbles and Filling the Reference
Sensor in Section 3.
Check that the sensors are installed in the correct order and are aligned, the
O-rings are in place, and the spring-loaded latch is closed as described in
Removing and Checking the Sensors on page 4-65.
Check the sensor contacts and measurement block for salt buildup. Check
behind the sensors for fluid leaks that can come from the reference sensor. If
salt or leaks are present, clean the sensor with reagent water, dry, and perform a
two-point calibration after reinstalling the sensor.
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......