The hemolyzer uses ultrasonic sound vibrations to rupture red blood cell
membranes and release hemoglobin molecules. These vibrations also rupture other
cells in the sample, reducing the light scattering from whole cells that can interfere
with hemoglobin analysis in the sample chamber. All CO-ox module samples pass
through the hemolyzer before entering the sample chamber.
The electronic components direct the operation of all system components. The
electronic components communicate with fluidic, measurement, and CO-ox
module components, and with external devices. The electronic components obtain
information from the sensors and perform all calculations required to determine
reported parameters. Table 1-2 describes these components and their functions.
Power Supply
Converts the line voltage from AC to DC and then
provides power to the rest of the system
Real Time
Processor (RTP)
Sends commands to other PC boards to perform system
activities; the RTP controls the fluidic system, acquires
data from the measurement module, calculates parameters,
and communicates with the user-interface processor
User-Interface Processor
(UIP)
Controls user interaction with the screen, the bar code
scanner, the printer, the database, and the external
communication ports
Backplane
Distributes power to each of the attached PC boards; the
backplane connectors route signals to electronic and
electromechanical subsystems
Control Board
Contains all of the drivers to control the fluidic system
valves and pumps, the preheater, and the sample port
Data Acquisition System
Board (DAS)
Performs the analog to digital conversion for a selected
measurement channel; after the conversion, the data is sent
to the RTP PC board for any required calculations
Printer Control
Receives information from the UIP and directs the roll
printer to print reports
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......