![Bayer HealthCare Rapidlab 800 Operator'S Manual Download Page 299](http://html1.mh-extra.com/html/bayer-healthcare/rapidlab-800/rapidlab-800_operators-manual_2712860299.webp)
Sensor drift is beyond predefined limits during a one-point or a two-point calibration.
Qualifiers:
Glu
Lac
If D codes also exist for reagent problems such as D23, D24, D29, or
D50, correct these problems first, then correct this D code.
Check the expiration dates and the levels of the reagents. Ensure that the
reagent bottles are installed properly and in the correct location. If reagents
have expired, are low, or are incorrectly installed, replace the affected reagents,
and perform a two-point calibration.
Check that the biosensors are installed in the correct order and are aligned, the
O-rings are in place, and the spring-loaded latch is closed as described in
Removing and Checking the Sensors on page 4-65.
Check the biosensor contacts and measurement block for salt buildup. Check
behind the biosensor for fluid leaks that can come from the reference sensor. If
salt or leaks are present, clean the sensors with reagent water, dry, and perform
a two-point calibration after reinstalling the biosensors.
Perform the Measurement Test for the affected biosensor as described in
Measurement Test, page 4-70.
Check the diverter valve to verify that the valve is working. Perform the Valve
Test as described on page 4-62.
If the glucose or lactate biosensors have been on the system less than 1 day, the
sensors may require additional time to hydrate the membranes.
Replace the glucose or lactate biosensor as described in Replacing the Glucose
and Lactate Biosensors in Section 3.
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......