Out of Range message
The result is above (up arrows) or below (down arrows)
the limits of the system’s measurement range, as described
in Appendix E, Performance Characteristics.
1.
Check the source and result, and ascertain whether a
value outside the range is likely.
2.
Check for sample movement during measurement.
3.
Check the fluid in the sample path for bubbles.
4.
Perform a wash.
5.
Ensure that you use the correct sample collection
techniques and anticoagulant, as described in Sample
Collection Devices and Anticoagulants in Section 1.
6.
Ensure that you use the recommended storage,
handling, and mixing techniques.
7.
Retest the sample.
↑
or
↓
↑↑
or
↓↓
The result is above (up arrow) or below (down arrow) the
established range.
1.
Check the sample source and result, and ascertain
whether a value outside the range is likely.
2.
Check for sample movement during measurement.
3.
Check the fluid in the sample path for bubbles.
4.
Perform a wash.
5.
Ensure that you use the correct sample collection
techniques and anticoagulant, as described in Sample
Collection Devices and Anticoagulants in Section 1.
6.
Ensure that you use the recommended storage,
handling, and mixing techniques.
7.
Retest the sample.
# is printed next to the
glucose or lactate result on
reports
The system detects substances in the sample that may
interfere with glucose or lactate measurement.
1.
Check the sample source and determine whether the
presence of an interfering substance is likely. See
Appendix E, Performance Characteristics, for a list of
interfering substances.
2.
If the message appears on multiple patient samples
that do not contain interfering substances, replace the
appropriate biosensor, as described in Replacing the
Glucose and Lactate Biosensors in Section 3.
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......