NOTE:
If your system has a CO-ox module, follow the procedure described for
the appropriate base model. For example, information identified for an 860 also
applies to an 865.
1.
Access the Condition screen from the Menu screen:
a. Select
2 Maintenance
and press
Enter
.
b. Select
2 Condition
and press
Enter
.
CAUTION:
Do not expose the glucose and lactate biosensors to the conditioner.
Replace the biosensors with the test/blank sensors (TB4) before conditioning.
Reinstall the biosensors within 2 hours.
CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
2.
Take the appropriate action.
840 or 850
go to step 3.
860
a. Remove the glucose and lactate biosensors.
b. Install the test/blank sensors.
c. Go to step 3.
3.
Draw the conditioner solution into a syringe.
4.
Insert the syringe into the sample port.
5.
Press
Analyze
.
6.
When prompted, remove the syringe.
7.
Wait 10 minutes for the conditioning cycle to finish.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
NOTE:
As you reinstall the biosensors, ensure that the biosensors are in the
correct location. Visually verify that you align the contacts on the biosensors with
the contacts in the measurement module. You can slide the remaining sensors to
the right to create more space.
8.
Remove the glucose and lactate test/blank sensors and reinstall the glucose and
lactate biosensors as described in
Replacing the Glucose and Lactate
Biosensors
, page 3-86.
2
Menu Code
2
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......