This procedure refers to right and left as you are looking at the rear of system.
NOTE:
Placing a piece of paper under the CO-ox module will allow for easier
movement on the work surface.
1.
Move the CO-ox module close to the right side of the 800 base model.
2.
Ensure that the CO-ox module hydraulic wall is open and slide the hydraulic
wall into position.
3.
Install the CO-ox power cable:
a. Insert the CO-ox module power cable through the access hole in the side of
the 800 base model.
b. Gently pull the cable through the access hole while pushing the CO-ox
module as close to the 800 base model as possible.
c. Connect the CO-ox module power cable to J38 on the 800 base model
Backplane board.
4.
Install the side panel from the base model to the CO-ox module.
5.
Close the 800 base model, but do not tighten the locking screws.
6.
Lift the side of the 800 base model up slightly to ensure that the lip of the
CO-ox module is seated in the groove located under the chassis shelf of the
800 base model.
Ensure that the two bosses on the CO-ox module are seated in the two
receptacles of the base model and that the screws are aligned with the bosses.
Do not tighten the screws.
7.
Move the CO-ox module hydraulic wall laterally along its hinge pin until the
sides of both units touch.
The two alignment bosses on the left side of the CO-ox module
hydraulic wall must be seated properly in the hydraulic wall receptacles of the
800 base model.
Do not crimp or pinch tubing between the hydraulic walls.
8.
Secure the hydraulic walls together:
Install the top rear screw first. You may need to remove the filter if it
prevents access to the top (upper rear) through hole. Be careful not to cut the
optics cable when securing the bottom (lower front) screw.
a. Ensure that the rear walls of both chassis are properly aligned.
b. Ensure that the captive screws are not engaged in the threaded portion of
the CO-ox hydraulic wall.
c. Insert a screwdriver into the access holes on the CO-ox module hydraulic
wall and tighten the screws to secure the hydraulic walls together.
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......