To minimize the errors these conditions can cause, use correct storage and handling
techniques. You can minimize errors due to metabolic changes by analyzing
samples as soon as possible after collection. This is particularly important for pO
2,
glucose, and lactate values, because the sample consumes oxygen and glucose, and
lactate is rapidly formed during storage. Lactate is produced by glycolysis and will
increase while the sample is stored. Glycolysis is temperature dependent. Lactate
increases approximately 0.1mM/hour at 4
°
C and 1.0 mM/hour at 37
°
C.* The rate
of oxygen consumption depends on several factors:
storage temperature
white blood cell count
reticulocyte count
If you cannot analyze samples within 10 minutes of collection, it may be
appropriate to place the samples in an ice-water slurry, with the portion of the
syringe containing the sample in contact with the ice-water slurry. You can store
samples in an ice-water slurry for up to 2 hours without significant change in
values for pH and pCO
2
; however this will affect the K
+
values. Samples with
elevated white blood cell or reticulocyte counts deteriorate more rapidly, and you
should analyze them immediately.
You can minimize errors due to room-air contamination by preventing air from
getting into the syringe. Immediately after drawing the sample, expel all the air
from the syringe, securely cap the syringe, and thoroughly mix the sample. Never
use corks as capping devices.
You can minimize errors due to incorrect sample mixing before analysis as
follows:
Roll a syringe between your hands and then gently invert the syringe several
times.
Gently invert a vacuum tube until the sample is homogenous.
Thoroughly mix a capillary sample until it is homogenous.
Blood cells settle during storage. If the sample is not well mixed before analysis,
the results obtained can be erroneous.
* Wandrup, Clinical Chemistry, 35/8, 1741, (1989)
.
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......