![Bayer HealthCare Rapidlab 800 Operator'S Manual Download Page 600](http://html1.mh-extra.com/html/bayer-healthcare/rapidlab-800/rapidlab-800_operators-manual_2712860600.webp)
21. Install the reagents:
Do not tighten or remove the bottle cap or attempt to mix the contents
of one bottle with another. The bottle caps are adjusted to ensure proper
reagent flow.
a. Remove the plugs from the reagent bottle caps.
b. Record the installation date in the space provided on the reagent labels.
c. Slide the bottles into position in the reagent compartment. The reagents
must be placed as follows from left to right on the reagent manifold:
22. Prepare the waste bottle:
a. Remove the waste bottle.
b. Remove the protective packaging from the waste detector.
c. Replace the waste bottle and ensure it is seated properly.
23. Prepare the reference sensor:
a. Partially fill the internal reference electrode compartment with 4M KCl
solution, and tap to remove bubbles.
b. Continue to fill the internal reference sensor compartment until the 4M KCl
fill solution flows into the reservoir.
c. Use the hex tool to remove the internal reference electrode from its
container. Insert the internal reference electrode into the reference electrode
compartment. Screw it into place using the hex tool. Do not cross thread
the internal sensor into the compartment. Tap the sensor to remove bubbles.
d. Remove the reservoir cap with the hex tool.
e. Fill the reservoir to the fill line with 4M KCl fill solution.
f. Put the reservoir cap back into position. Hand tighten the reservoir cap.
Do not overtighten. Overtightening can deform the gasket and cause
leakage.
g. Wipe any excess KCl solution from the exterior of the reference sensor
with a lint-free tissue.
h. Tap the sensor with your knuckle to remove bubbles.
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......