4.
Disconnect the tubing from the gas regulator and discard the tubing.
5.
Connect one end of the new tubing to the fitting on the regulator.
6.
Connect the other end of the tubing to the fitting on the reagent manifold.
NOTE:
The average main tank pressure is 2200 psi. The average secondary valve
pressure is 3 to 5 psi.
7.
Slowly open the gas tank by turning the valve stem counterclockwise with the
wrench.
Turn the valve stem approximately 3/4 turn until the regulator pressure gauge
indicator stops rising.
8.
Open the valve stem one more turn.
9.
Listen carefully for any gas leaks.
10. Check for leaks using soapy water and watching for bubbles.
11. Verify that the regulator outlet gauge indicates 3 to 5 psi.
12. Press
Continue
.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
13. Press
No
.
14. Check the gas flow rate from the Menu screen:
a. Select
3 Troubleshooting
and press
Enter
.
b. Select
1 Fluidics System
and press
Enter
.
c. Select
4 Valves
and press
Enter
.
d. Select the gas type that you replaced and press
Enter
.
e. Insert an aspiration adapter into the sample port and immerse the open end
of the adapter into a small container of reagent water.
f. Press
Start Test
.
g. Verify that a steady stream of bubbles flows into the water.
h. Press
Stop Test
.
i. Verify that the bubbles stop flowing into the water.
j. Remove the adapter.
k. Press
Exit Test
.
15. Perform a two-point calibration from the Menu screen:
a. Select
1 Calibration
and press
Enter
.
b. Select
4 Gas Two-point
and press
Enter
.
4
1
Menu Code
3
(from the Main Menu)
1
Menu Code
4
(from the Main Menu)
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......