The pCO
2
sensor is based upon the electrode described by Severinghaus and
Bradley.
15
It is a complete electrochemical cell that consists of a measuring
electrode and an internal reference electrode. The measuring electrode, which is a
pH electrode, is surrounded by a chloride bicarbonate solution. A membrane
permeable to gaseous CO
2
separates this solution from the sample. The internal
reference electrode, which contains a silver/silver chloride electrode surrounded by
the chloride-bicarbonate solution, provides a fixed potential.
Sample Path
Measuring Electrode
Contact
Internal Reference
Electrode Contact
As the sample comes in contact with the membrane, CO
2
diffuses into
the chloride-bicarbonate solution, which causes a change in the hydrogen ion
activity.
CO
2
+ H
2
O
HCO
3
–
+ H
+
The internal pH electrode detects the change in hydrogen concentration occurring
in the chloride bicarbonate solution and generates a half-cell potential. This
potential, when compared to the fixed potential of the reference electrode, results
in a measurement that reflects pH change in the chloride bicarbonate solution. The
change in pH is related to the log of the partial pressure of CO
2
.
Oxygen (O
2
) is essential for cell and tissue metabolism in the body. The
cardiopulmonary system is responsible for transporting oxygen to the cells.
Oxygen transport involves four major steps: convection and diffusion from the air
into the pulmonary circulation, combination of O
2
from the lungs with hemoglobin
in red blood cells, transportation of the O
2
through the arteries to the cell, and
finally the release into the tissues and utilization of O
2
at the cellular level.
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......