The CO-oximeter (CO-ox) module detects and indicates concentrations of
sulfhemoglobin greater than 1.5%.
Hemoglobin derivatives have characteristic absorbance spectra; that is, each
derivative absorbs light differently at different wavelengths. Similarly, interfering
substances also absorb light at known wavelengths.
The spectral absorption method determines concentration using matrix equations.
For each substance or fraction, the absorbance at a specific wavelength is equal to
the product of the path length, concentration of the fraction or substance, and the
molar absorptivity or the extinction coefficient for that substance, as shown in the
following equation:
A
x
ε
1
C
1
2
C
2
. . .
n
C
n
where A
x
is the absorbance at a specific wavelength,
ε
is the major extinction
coefficient for that fraction or substance at a specific wavelength, and C is the
concentration of the substance.
These equations are based on the work of VanAssendelft
33,34
and Benesch,
Benesch, and Yung.
35
The 800 systems CO-oximeter (CO-ox) module measures the light of whole blood
at several wavelengths. Based on this, the CO-ox module measures and reports
total hemoglobin and other related quantities. The CO-ox module also detects the
presence of interfering substances such as bilirubin, cyanmethemoglobin, turbidity,
and dyes.
The CO-ox module is connected to the base model, which supplies power to the
module. The CO-ox module contains measuring, fluidic, and electronic
components. It uses the sample entry port, reagents, and waste components of the
base model.
The measurement system for the CO-ox module detects and quantitates the
analytes present in the sample. The measurement system has the following
components:
the lamp
the illumination optics (lenses and filters)
the fiber optic coupler
the sample chamber
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......