Fluid detector 5 (FD5), the CO-ox sample chamber, or both do not detect reagent.
Qualifiers:
7.3
Wash
Check the expiration dates and levels of the 7.3/COox Zero and the Wash/Glu
Zero. Ensure that the reagent bottles are installed properly and in the correct
location. If reagents have expired, are low, or are incorrectly installed, replace
the affected reagent. Perform a one-point calibration if the D78 disappears.
Ensure that the CO-ox sample path is connected to the base model. If the
sample connector needs to be inverted, refer to Disconnecting the CO-ox
Sample Path, page 4-92 for instructions.
Check the CO-ox sample path for obstructions. If you see any obstructions,
remove obstructions as described in Removing Obstructions from the CO-ox
Sample Path, page 4-87.
Check the CO-ox sample tubing for leaks or cracks. If you find cracks or leaks,
replace the tubing as described in Replacing the CO-ox Sample Tubing in
Section 3.
Check the gasket in the hemolyzer for cracks and that the gasket is seated
correctly. If necessary, replace or reinstall the gasket as described in Cleaning
the Hemolyzer in Section 3.
Check the CO-ox sample chamber for obstructions and that the gasket is seated
correctly. If necessary, remove obstructions or reinstall the gasket as described
in Cleaning the CO-ox Sample Chamber in Section 3.
Perform a one-point calibration.
Perform the Fluidics Functions Test for the appropriate reagent as described in
Fluidics Functions Test, page 4-56.
Check the CO-ox pump tubing for leaks, flattened areas, or other signs of wear.
Replace the tubing, if necessary, as described in Replacing the CO-ox Pump
Tubing in Section 3.
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......