The wash did not completely clean the sample path due to low volume or poorly segmented flow.
If a D23 or D24 code also exists, correct these problems first, then
correct the D29.
Perform a wash.
Ensure that the bottle is installed properly. Check the level of Wash/Zero
reagent. If the level is low, replace the reagent as described in Replacing the
Reagent Bottles in Section 3.
Check for cracks or leaks in the sample tubing. If you find cracks or leaks,
replace the sample tubing, as described in Replacing the Sample Tubing in
Section 3.
Check the sample entry components for obstructions and remove them as
described in Removing Obstructions from the Sample Entry Components,
page 4-79.
Check the measurement module for leaks. Check that the sensors are installed
in the correct order and are aligned, the O-rings are in place, and the
spring-loaded latch is closed as described as described in Removing and
Checking the Sensors on page 4-65.
If you see obstructions, remove them as described in Removing Obstructions
from the Measurement Module, page 4-82.
Perform the Valves Test as described on page 4-62 to verify that the wash,
diverter, vent, and wash bypass valves are functioning.
Perform the Pump Functions Test as described on page 4-58.
Check the pump tubing for leaks, flattened areas, or other signs of wear. If
necessary, replace the tubing as described in Replacing the Pump Tubing in
Section 3.
If the D code reappears, contact your Service Representative.
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......