Use wash reagent that has not expired and has not been open longer
than 60 days. Do not install the wash reagent bottle that you use for this
procedure on the system.
d. If the obstruction is not removed, fill the syringe with fresh wash reagent
and inject the wash reagent into the sensor sample path.
e. Dry thoroughly and reinstall the gas sensors, ensuring that the O-rings are
in place.
To avoid damage to the sensors, use the clot-removal line
carefully. Be particularly careful with the pH sensor.
4.
Inspect the reference sensor, the sample ground/temperature sensor, pH sensor,
and the Na
+
sensor for obstructions:
a. Remove the sensors one at a time and inspect the sample path for
obstructions.
b. Cut a piece of tubing, or use a capillary adapter, and place the tubing on the
tip of the syringe.
c. Push a 0.558 mm (0.022-inch) diameter clot-removal line through the
sensor to remove obstructions, as shown in Figure 4-17.
$!'& % "#!$#" !
! !
!$ !#$! "!
Push the spring-loaded
latch to the right.
Lift the sensor up
and out of the
measurement module.
Clot-removal
Line
d. If the obstruction is not removed, cut a piece of tubing, or use a capillary
adapter, and place the tubing on the tip of the syringe.
e. Place the syringe against the sensor sample path and inject air through the
sensor to remove obstructions.
f. If the obstruction is not removed, fill the syringe with fresh wash reagent
and inject the wash reagent into the sensor sample path.
Summary of Contents for Rapidlab 800
Page 1: ...2000 Bayer Corporation All rights reserved 115701 Rev C 4 2000...
Page 13: ......
Page 14: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ...3 3 0 3 3 3 3 1 3 1 3 3 3 3 3 12 3 3 1 1 3 1 3 3 1 3 3 12 3 3 3 1 1 3 1 3 3...
Page 25: ......
Page 26: ......
Page 33: ...CO ox Pump CO ox Module Sample Chamber Hemolyzer...
Page 34: ...Power Input Fuse Compartment Diskette Eject Button Diskette Drive Air Filter Cover...
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ...0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0...
Page 100: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ...2 0 5 5 5 0 5 0 5 5 34 5 4 5 5 0 5 0 1 5 2 0 5 5 2 5 3 5 0 5 0 5...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 213: ......
Page 214: ......
Page 235: ......
Page 236: ......
Page 280: ......
Page 281: ......
Page 282: ......
Page 283: ...2 1 2 1 2 1 2 1 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 0 2 2 2 2...
Page 285: ......
Page 286: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 339: ......
Page 340: ......
Page 364: ......
Page 365: ......
Page 366: ......
Page 409: ......
Page 410: ......
Page 496: ......
Page 497: ......
Page 498: ......
Page 578: ......
Page 620: ......
Page 660: ......
Page 661: ......
Page 662: ......
Page 663: ......
Page 664: ......
Page 665: ......
Page 666: ......
Page 671: ......
Page 672: ......
Page 687: ......
Page 688: ......