2 Menu Item Selection
Reflex HC Series
45
Load the sample when the value displayed is within the Zero
Range. When the stability symbol has illuminated and the
displayed positive or negative value is more than 5 times the
Zero Range, data is automatically output.
Unload the sample or press the
key. When the
displayed value is within the Zero Range and the stability
symbol ha illuminated, data is output again.
Menu Map item [46]
(on /continue)
By pressing the
key while the
and
symbols
are lit, the
symbol goes out, the
symbol lights and
the displayed data is continuously output.
Continuous output stops temporarily when the
key is
pressed.
Note:
During continuous output, the
symbol may appear to
remain lit. If the transfer speed of the data output is slow, the
display may flash. Increase the transfer speed as much as
possible and set the handshake to
.
Menu Map item [47]
(on / go)
When the graphic display is in Comparator mode (Group mode
or Level mode) and the stability symbol lights in the GO range,
the data is output once.
The next data output is executed after the displayed value is
within the Zero Range.
2.8.5 Peak
Hold
Menu Map item [49] measures the displayed peak value. The
symbol is illuminated when the Peak Hold function is
activated.
"Peak value" is the highest or lowest stable value displayed
after the display has changed beyond five times the Zero
Range.
(1)
In the peak detection standby state with the and
symbols illuminated, press the
key to
tare the display.