1 Basic Operation
Reflex HC Series
17
Note:
If ENVIRONMENT menu 37
has
been selected, the display stops in the whole lighting
state. In this case, press the
key during whole
lighting to change the display to the kilogram-display.
1.4
Adjusting the Built-in Clock
The Reflex HC series of balances have a built-in battery driven
clock. After unpacking and installing the balance, adjust this
clock to the appropriate date and time.
See Section 2.10.1 Date for the Built-in Clock
See Section 2.10.2 Time for the Built-in Clock
1.5 Span
Calibration
It is necessary to calibrate the balance after it is moved, or
when the ambient temperature has changed.
Verify that the balance is stable before performing the span
calibration. To achieve a very stable state, ensure that the
balance has been turned on for at least one hour, that the
temperature is constant, that there are no breezes or
vibrations and that the balance is in an area isolated from the
normal traffic flow.
Refer to Section 2.5.2 “Span Calibration Using the External
Weights” for more information on the Span Calibration
procedure.
1.6 Weighing
(1)
Place the weigh vessel (tare) on the pan.
(2) Press
the
key to zero the display.
Note:
If a Pretare value is set, the
symbol is
illuminated and zeroing of the display does not occur.
(3)
Place the object to be weighed on the pan, and read the
displayed value after the stability symbol is displayed.
Errors Displayed During Weighing
Overload: Weighing capacity has been exceeded.