2 Menu Item Selection
Reflex HC Series
36
(1)
Load the indicated calibration mass and press the
key.
(2)
The zero display flashes.
(3)
Unload the weight from the pan and press the
key.
(4)
The display changes to the xxx display. (xxx indicates
a numeric value)
(5)
To change the value to zero, press the
key.
(6) Press
the
key to avoid changing the value to
zero. Pressing the
key interrupts calibration and
does not change this value to zero.
Note:
Changing the value to zero is equivalent to
performing span calibration.
2.6 Graphic
Display
Menu
:
2.6.1 Overview
The Reflex HC series balance has a bar graph located on the
left side of the display. This graph can be conveniently used
for liquid weighing and pass or failure judgment (comparator
function) of the sample.
The graphic display functions include the Full Scale mode,
Target mode, Group mode, Level mode, and no graphic
display mode.
Note:
Target, limit, upper, and lower values are set as numeric
values only. Set the correct numeric value for the unit that will
be used for weighing.
y
In weighing mode, changing the displayed unit does
NOT change the target, limit, upper, or lower numeric
values. For example, when the upper limit has been set
at 10g, switching the unit from “g” to “kg” does not
change the limit to 10kg. It will be 0.01kg.
y
The decimal point is invisible in the value setting display.
Determine its position based on the resolution of the
balance. For example, if the balance’s minimum display