2 Menu Item Selection
Reflex HC Series
34
Span Calibration:
Adjustment of the balance to specifications
using two weight values; zero and an
appropriate value for the balance capacity.
Calibration Check:
Comparing the current calibration mass
reading to the calibration mass reading after
the last span calibration.
Factory Default Settings
Reflex HC series
Calibration using the external weight.
(
)
The settings for the type of calibration can be changed with
menu selection.
Section 2.6.1 "Selecting Standard Calibration Type."
Note:
The following displays may appear in the situations
listed.
When it takes too much time to stabilize the
balance.
When the zero point of the balance appears to
have changed greatly.
When the balance calibration appears to have
changed greatly or the wrong weight is loaded on the pan
during calibration.
In all of these situations, span calibration or calibration check
is not performed.
Refer to Section 4.4.2 “Error Display.
2.5.2
Span Calibration Using External Weights
(1)
Verify that the balance is in mass display and unload the
sample from the pan.
(2) Press
the key once.
is displayed.
(3) (If
is not displayed, return to mass display
and select Menu Map item [3].)
Press the
key.
The value of the correct calibration mass to be loaded is
displayed and flashes.